Calibrating and Configuring the Sensor using the H

2

Configuration Utility Software

The simplest way to set up the analog outputs, calibrate the analog outputs, and calibrate the

Hydrogen sensor is to use the SSi H

2

Configuration Utility software. This software provides a

simple to use graphical interface for performing the setup and calibration functions. It

communicates via Modbus (either RS-232 or RS-485). This is set by moving dip switch #4 to the

off position. To communicate via RS-232, use the 9-pin connector with the jumpers in positions

2 and 3. To communicate via RS-485, use the RT+ and RT- terminals on the 8 pin terminal strip.

Minimum Computer Requirements
• Microsoft Windows XP/Vista/7
• 500 MHz CPU
• 128MB RAM
• 2MB hard disk storage space
• 1 RS-232 or RS-485 Serial Port

Configuring Communications

Open the H2 Configurator and click Options  Settings to open the dialog that allows you to set

the serial port, baud rate and target address of the H2 cell. (See Figure 8.)

Figure 8 - Settings window

When using the RS-232 port, the baud rate should be set to 9600 and the address should be set

to 250. 250 is an SSi broadcast address and any H

2

cell that sees the message will answer, so

while the 250 address should be used for RS-232 it should not be used for RS-485 if there are

multiple SSi instruments on the same serial port. The Port setting may need to be adjusted to

match the port on your PC that is connected to the H

2

cell. RS-232 communications use a

simple straight through cable to the DB9 connector.

Overview Screen

The Overview screen (Figure 9) displays the current % H

2

and the three buttons used to

configure or calibrate the H2 cell.