Plumbing connections, Analog output assignments, Plumbing connections analog output assignments – Super Systems H2 Sensor User Manual

Page 8

Advertising
background image

Hydrogen (H

2

) Sensor Operations Manual

Super Systems Inc.

Page 8 of 17

Applies to Loop 1 (#1 and COMMON)

Applies to Loop 2 (#2 and COMMON)

JP2

JP3

Resulting

Output

JP4

JP5

Resulting

Output

Open

Open

4-20mA

(current)

Open

Open

4-20mA

(current)

Closed

Open

2-10 VDC

Closed

Open

2-10 VDC

Open

Closed

4-20mA

(current)

Open

Closed

4-20mA

(current)

Closed

Closed

1-5 VDC

Closed

Closed

1-5 VDC

IMPORTANT: If Modbus register 43 is set to 0,

the above conditions are true. If Modbus register

43 is set to 1, the following changes take place

(depending on jumper settings):
• 4-20mA becomes 0-20mA
• 2-10 VDC becomes 0-10 VDC
• 1-5 VDC becomes 0-5 VDC.

IMPORTANT: If Modbus register 44 is set to 0,

the above conditions are true. If Modbus register

44 is set to 1, the following changes take place

(depending on jumper settings):
• 4-20mA becomes 0-20mA
• 2-10 VDC becomes 0-10 VDC
• 1-5 VDC becomes 0-5 VDC.

Table 1 - Analog Output Jumper Settings

Plumbing Connections

(Items “G” and “H” in Figure 2)

The sensor is provided with barb fittings (Items “G” and “H”) that are intended for use with 1/8”

ID flexible tubing. Flow goes into “G” and out through “H”. The barb fittings can be removed at

the user’s discretion, and any fitting with a 1/8” male pipe thread can be used in their place. If

the fittings are going to be replaced with different fittings, they must be replaced while the

stainless steel sensor block is still attached to the circuit board. Do not remove the sensor

block since its alignment with the sensors is a critical function that should only be performed by

a trained technician.

Analog Output Assignments

There are two analog outputs on the sensor. The default settings result in Output 1 being set

for Percent H2 (0-100%) and Output 2 being set for Percent Dissociation (0-100%). Both of

these outputs can be configured for any of the following parameters:

Percent Hydrogen (H

2

)

Percent Dissociation (DA)

Percent Ammonia (NH

3

)

Nitriding Potential (K

N

)

External

In “External” output mode, no calculation is performed, and the output is set to match a specific

value. Modifications to the Analog Output Assignments can be performed with the SSI H

2

Cell

Configuration Utility Software, which is described in the next section of this manual.