Appendix 1: modbus register map – Super Systems H2 Sensor User Manual

Page 14

Advertising
background image

Hydrogen (H

2

) Sensor Operations Manual

Super Systems Inc.

Page 14 of 17

Appendix 1: Modbus Register Map

Register Functions and Default Values

#

Function

Default

0 Current firmware revision level

Varies

1 RS-232 communication mode (0=Modbus slave, 1=ASCII /Previous SSI

Sensor)

DIP Sw.

2 RS-232 baud rate (0=9600, 1=19200, 2=38400)

0

3 RS-485 communication mode (0=Modbus slave, 1=Modbus master [NOT

USED])

0

4 RS-485 baud rate (0=9600, 1=19200, 2=38400)

1

5 Temperature trim for thermistor 1 / ambient temperature

200

6 A/d counts for thermistor 1

Varies

7 Thermistor 1 temperature

Varies

8 %H2 A/D counts – High (Right justified: 0x00 byte, high byte, mid byte,

low byte)

Varies

9 %H2 A/D counts – Low (Right justified: 0x00 byte, high byte, mid byte, low

byte)

Varies

10 %H2 measured voltage

Varies

11 %H2 as found from a polynomial

Varies

12 %H2 sensor calibration (0=none, 1=zero cal, 2=span cal)

0

13 %H2 gas composition (value/100 = %H2/N2 balance)

0

14 Sensor Modbus address (important for slave communications only)

DIP Sw.

15 Model Number

4100

16 Set Factory Defaults (23205=Full, 23206=%H2, 23207=Loop 1, 23208 =

Loop 2)

0

17 4-20mA #1 DAC output counts

Varies

18 4-20mA #2 DAC output counts

Varies

19 4-20mA #1 Source (0=%H2, 1=%DA, 2=%NH3, 3=Kn, 4= External)

0

20 4-20mA #1 Zero value

Varies

21 4-20mA #1 Span Value

10000

22 4-20mA #1 External current loop (Used only if register 19 = 4)

0

23 4-20mA #2 Source (0=%H2, 1=%DA, 2=%NH3, 3=Kn, 4= External)

1

24 4-20mA #2 Zero value

0

25 4-20mA #2 Span Value

1000

26 4-20mA #2 External current loop (Used only if register 23 = 4)

0

27 Enable Calibration of 4-20ma output (0=disable, 1=enable)

0

28 Calibrate 4-20mA output (1=#1 zero, 2=#1 span, 3=#2 zero, 4=#2 span)

0

29 4-20mA calibration value (4000 to 20000 for 4-20mA)

0

30 Calibration result codes (9-12=successes, 13-16=failures, 20=no cal

specified)

0

31 Percent DA value

Varies

32 Percent NH

3

value

Varies

33 Super Kn value

Varies

34 %N

2

flow

0

35 %NH

3

flow

0

36 %DA flow

0