Super Systems 7SL User Manual

Page 33

Advertising
background image

3 3

AL

Current Input Initial Scale Value

a) Connect calibrator and instrument as shown in

Figure 7.

10

9

Figure 7. Calibrator Connection

b) The upper display shows “OFF”, the lower displays

shows “AL”.

c) Set calibrator to 0.000 mA (even if the minimum range

value is 4mA).

d) Push the button; the display changes to “ON”.
e) After a few seconds, start calibration by pushing the

FUNC button. When this calibration is complete, the
instrument will go to the next parameter.

AH

Current Input Final Scale Value

a) Set the calibrator to 20 mA
b) Push the button; the displays will show “ON”

and “AH”.

c) After a few seconds, start calibration by pressing the

FUNC button. When this calibration is complete, the
instrument will go to the next parameter.

A.

Current Input Check

The display shows “A.” followed by a number showing the
measured value in counts.

a) Check the calibration (linear) by setting:

0.000 mA – the readout must be equal to
“A.000 00” ± 10 counts;
20.000 mA – the readout must be equal to
“A.300 00” ± 10 counts;
10.000 mA – the readout must be equal to
“A.150 00” ± 10 counts.

b) Push the FUNC button to go to the next parameter.

nL

5 VOLT Input Initial Scale Value

a) Connect calibrator and instrument as shown in

Figure 8.

10

9

Figure 8. Calibrator Connection

b) The upper display shows “OFF”, the lower displays

shows “nL”.

c) Set calibrator to 0.000 V (even if the minimum range

value is 1V).

d) Push the button; the display changes to “ON”.
e) After a few seconds, start calibration by pushing the

FUNC button. When this calibration is complete, the
instrument will go to the next parameter.

nH

5 Volt Input Final Scale Value

a) Set the calibrator to 5.000 V.
b) Push the button; the displays will show “ON”

and “nH”.

c) After a few seconds, start calibration by pressing the

FUNC button. When this calibration is complete, the
instrument will go to the next parameter.

n.

5 Volt Input Check

The display shows “n.” followed by a number showing the
measured value in counts.

a) Check the calibration by setting:

0.000 V – the readout must be equal to
“n.000 00” ± 10 counts;
5.000 V – the readout must be equal to
“n.300 00” ± 10 counts;
2.500 V – the readout must be equal to
“n.150 00” ± 10 counts.

b) Push the FUNC button to go to the next parameter.

7sl-6-00.p65

9/17/02, 2:44 PM

33

Advertising