9 ELECTRICAL

SPECIFICATIONS 4-wire

system

Output functions

continuous level measurement through analog output and
point level detection through switching output

Analog output (active)

current output 4…20 mA
the span between the lower range value [4mA] and the upper range value
[20mA] is equal to 0...100% of the continuous level measurement reading. It
is recommended that the span between those two range values stays within
the measuring range [M] (see figure 10)

Total load resistance

<500 Ω: HART

TM

resistor approx. 250Ω + load resistance approx. 250Ω

if the current output is connected to a device with an inner resistance of ap-
prox. 250 Ω, then there is no additional, external HART resistor necessary. In
that case, the HART modem is connected in parallel to the current output
wires

Lower range value

4,0 mA (span 0%)

Upper range value

20,0 mA (span 100%)

Response time

0,5 s (default), 2 s, 5 s (selectable)

Temperature drift

<0,2 mm/K change in ambient temperature

Switching output DC PNP (active) NC or NO (short-circuit protected)

Load

current <200mA

Signal voltage HIGH

supply voltage - 2 V

Signal voltage LOW

0 V…1 V

Response

time <200

ms

Supply voltage

12…30 VDC (reverse-polarity protected)

Current consumption

<70 mA at 24 VDC (no burden)

Start-up time

<6 s

Cable terminals

screwless, cage clamp terminal block for stranded and
solid wires 0,5…2 mm²
the usage of cable end sleeves with insulation collar is not recommended

10 APPLICATION

SPECIFICATIONS

continuous level measurement and point level

detection in liquids

Dielectric constant [ε

r

]

single rod probe and rope probe: >1,8

coaxial probe:>1,4

Conductivity no

restrictions

Density no

restrictions

Dynamic viscosity

single rod probe and rope probe: <5.000 mPa s = 5.000 cP
coaxial probe: <500 mPa s = 500 cP

Application temperature

-40°C…+150°C

Ambient temperature

operation: -25°C…+80°C storage: -40°C…+85°C

Application pressure

-1bar...40 bar

Velocity of level change

<1.000 mm/s

Interface (e.g. oil on top of water)

an oil layer of <70mm thickness on top of water is not detected by the sensor;
in this case the sensor will detect only the water level at a slightly lower posi-
tion than actual. From an oil layer thickness >70mm onwards, the sensor
detects the total level, including the oil layer, according to specifications