This task can be used on any K2 Summit system, but it is required on any system that has a Type
II (ADLINK) CPU carrier module. This includes the first generation K2 Summit system, which can
have a Type II CPU carrier module that was installed in the factory or that was upgraded in the field.

NOTE: This procedure is intended for use by Grass Valley Service personnel or under the direct
supervision of Grass Valley Service personnel.

After you replace a disk controller board, you must import the configuration information from the
existing disks. This allows the new board to see the LUNs as previously configured.

1. After replacing the disk controller board, power up the K2 Summit system.

Ignore SNFS messages that can open at any time during this procedure.

2. On the Windows desktop, open the

MegaRAID Storage Manager

icon.

3. When prompted, enter administrator credentials.

The MegaRAID Storage Manager main window opens.

4. In the MegaRAID Storage Manager main window tree-view, verify that drives are reported as

"Unconfigured Drives".

02 February 2012

K2 Summit 3G Production Client Service Manual

65

Service procedures