Maintenance – Amprobe AT-7000 User Manual

Page 34

Advertising
background image

30

Transmitter fuse replacement:

Figure 5.4: Transmitter fuse replacement

�

Warning:

To avoid shock, injury, or damage to the Transmitter, disconnect test

leads before opening case.

1. Disconnect all test leads from the Transmitter.
2. Make sure that the transmitter is turned off.
3. Use the star screw driver to unscrew the holding screws.
4. Remove the fuse cover by pulling it upwards as shown in figure 5.4.
5. Remove the fuse from the fuse holder.
6. Insert the new fuse in the fuse holder. 3.15A , 600V MAX, SLOW 5X20MM
7. Insert the fuse cover and secure it with the holding screws and tighten with the Star

screw driver.

5. maintenance