Orienting the sensor, Flow direction arrow, Vertical pipeline – Emerson CMF200A User Manual

Page 12: Step 2 orienting the sensor

Advertising
background image

10

Micro Motion

®

ELITE

®

Sensors

Orienting the Sensor

Step 2

Orienting the Sensor

The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid.
Micro Motion recommends orienting ELITE sensors as shown in Table 3.

Flow direction arrow

The sensor has a flow direction arrow (see Figures 1–4) to help you configure the transmitter for flow
direction. If possible, install the sensor so that the flow direction arrow matches actual process flow.

Vertical pipeline

If the sensor is installed in a vertical pipeline, liquids and slurries should flow upward through the sensor.
Gases may flow upward or downward.

Table 3

Recommended sensor orientations

Sensor model

Application

Recommended orientation

CMF010

All applications

Tubes flat
Horizontal pipeline

CMF025, CMF050,
and CMF100

Liquids

Tubes down
Horizontal pipeline

Gases and slurries

Tubes up
Horizontal pipeline

CMF200, CMF300,
CMF400, and
CMFHC3

Liquids

Tubes down
Horizontal pipeline

Gases

Tubes up
Horizontal pipeline

Slurries

Flag mount
Vertical pipeline