Installation Manual

13

Mounting the Sensor

Figure 10

Mounting the temperature extender onto the sensor

Mounting the electronics of high-temperature and extreme high-temperature sensors

High-temperature and extreme high-temperature ELITE sensors come with a 32

″ (812 mm) pre-installed

flexible conduit. This conduit is required for agency approval.

A factory-supplied Model 2400S transmitter, Model 1700/2700 transmitter, core processor, or junction box is
connected to the end of the flexible conduit. Mount the electronics to a wall or instrument pole using the
supplied bracket (Figure 11).

Extended mount — carefully align the notches at
the bottom of the extended mount with the
notches on the feedthrough to engage properly

Plastic plug — remove and
discard

Feedthrough — note the

notches for correct

engagement of the extender

Clamping ring screw —

when extender is engaged,

tighten this to 13–18 in-lbs

(1.5–2 N-m)

Model 2400S transmitter or enhanced core
processor — preinstalled on the extended
mount