Installing wafer-style sensors – Emerson CMF200A User Manual

Page 16

Advertising
background image

14

Micro Motion

®

ELITE

®

Sensors

Mounting the Sensor

Figure 11

Electronics mounting for high-temperature and extreme high-temperature
ELITE sensors

Installing wafer-style sensors

A wafer-style sensor, which has no flanges or fittings, lets you “clamp” the sensor between process
connections in the pipeline. Model CMF025, CMF050, and CMF100 sensors are available in the wafer style.

A wafer installation kit is shipped with a wafer-style sensor. A wafer kit contains the following pieces:

• 4 flange bolts

• 8 flange nuts

• 2 alignment rings, which help center the sensor between the bolts

To install a wafer-type sensor, refer to Figures 12 and 13 and follow the steps below:

1. Make sure the bolts provided in the wafer installation kit are rated for your process connection.

2. Slip the sensor alignment rings over each end of the sensor wafer, then insert the sensor between the

process connections in the pipeline (Figure 12). Installing gaskets is recommended. (Micro Motion
does not supply gaskets.)

3. Insert the bolts through both process connections, and thread the nuts onto the bolts. Tighten nuts as

tight as you can with your fingers.

4. Rotate the sensor alignment rings in the direction that pushes the bolts outward (Figure 13). Rotate

both rings until the assembly is centered and tight.

5. With a wrench, tighten nuts in an alternating order, to ensure the process connections are evenly

tightened.

Use 4 × 5/16

″ (8 mm)

bolts to secure the

bracket to a wall

Mounting

bracket

Use 2 user-supplied

5/16

″ (8 mm) U-bolts

to secure the bracket

to an instrument pole

Flexible
conduit

Mounting

bracket

Flexible
conduit

Transmitter, core processor, or
junction box (Model 2400S shown)

Advertising