Rupture disks – Emerson CMF200A User Manual

Page 24

Advertising
background image

22

Micro Motion

®

ELITE

®

Sensors

Rupture Disks

Rupture Disks

If the sensor has rupture disks, they are installed in the sensor purge fitting openings. The rupture disks should
remain installed at all times. If you remove a rupture disk from the sensor case, it will be necessary to re-purge
the case (see Case purging procedure on page 21).

Rupture disks are meant to vent process fluid from the sensor case in the event of a flow tube rupture. Some
users connect a pipeline to the rupture disk to help contain escaping process fluid.

Stay clear of the rupture disk pressure relief area. High-pressure fluid escaping from the sensor can
cause severe injury or death.

If a rupture disk is removed, the instructions for repurging the case provided under Case purging
procedure
on page 21 must be followed. Improperly pressurizing the sensor case can result in serious
injury.

For more information about rupture disks, contact Micro Motion Customer Service.

Table 5

Time required to purge ELITE sensor cases

Sensor model

Purge rate
ft

3

/hr (l/hr)

Time

(1)

minutes

(1) If purge lines are being used, increase purge time to fill the additional volume.

CMF010

20 (566)

1

CMF025

20 (566)

1

CMF050

20 (566)

2

CMF100

20 (566)

5

CMF200

20 (566)

12

CMF300

20 (566)

30

CMF400

20 (566)

55