12.6 Channel Menu

Each High-Level Input channel is configured and calibrated via its channel menu.
Detailed instructions on operating the instrument in the SETUP Menu mode can
be found in “SETUP Menu mode” on page 23. A diagram of all menus is located
in “Setup Menu Reference” on page 157.

12.6.1

OPERATION

Sub-Menu

This menu controls the operation of this channel when the instrument is in the
RUN mode.

AUTO-ZERO

Menu Item

This enables or disables the channel’s automatic zeroing function. The auto-zero
function can remove the effect of slight temperature shifts in the transducer’s out-
put. The two options for this menu item are “

ON

” and “

OFF

”.

When enabled, the channel will reset its tracking value to zero when the tracking
value has never been more positive than +10 or more negative than -10 display
counts for any 10 second period.

For example, assume that the

DISPLAY.DECPT

menu item is set to

000.00

(two deci-

mal points). If the tracking value never leaves the range of

-000.10

to

+000.10

for

any 10 second period, then tracking value will reset to

000.00

automatically.

FREQ. RESPONSE

Menu Item

This sets the frequency response, step response, and resolution of the channel
according to the table below. The tracking value, peak/valley detector values and
analog output on this channel are all affected. The default value of

016. HERTZ

is

suitable for most applications.

POWER-ON SOURCE

Menu Item

This menu selects which value is displayed by the channel when first entering the
RUN mode.

The choices are:

TRACK

” means the live tracking value of the channel.

NOTICE

To compensate for the transducer’s initial warm-up shift, the auto-zero function
operates every 2 seconds during the first 60 seconds of RUN mode operation.

Frequency

Response

Step Response

(ms) (typical)

Resolution (counts) (not including min.

10% overrange/underrange capability)

5V & 10V Inputs

1V & 4-20mA Inputs

002. HERTZ/FAST

40

±50000

±50000

002. HERTZ

440

±50000

±50000

008. HERTZ

110

±25000

±25000

016. HERTZ

55

±25000

±20000

032. HERTZ

28

±20000

±10000

050. HERTZ

16

±15000

±10000

100. HERTZ

8

±10000

±5000

250. HERTZ

3

±10000

±5000

500. HERTZ

2

±4000

±2500

800. HERTZ

2

±2500

±2500