Honeywell SC1000 User Manual

Page 90

Advertising
background image

page 90 008-0608-00

previously entered as

KNOWN POINT 3/5

and

DISPLAY. UNITS

). When you have applied

this displacement to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00750 UNIT

(where 00750 and UNIT are pre-

viously entered as

KNOWN POINT 4/5

and

DISPLAY. UNITS

). When you have applied this

displacement to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are pre-

viously entered as

KNOWN POINT 5/5

and

DISPLAY. UNITS

). When you have applied this

displacement to the transducer, press [ENTER].

Upon completion, the display will indicate

DONE

and the instrument will return to the

RUN mode.