KNOWN POINT 2/3

”:

point 2 of 3, usually 50% of the transducer’s full-scale capac-

ity.

KNOWN POINT 3/3

”:

point 3 of 3, usually 100% of the transducer’s full-scale

capacity.

When using the 5-Point Known Load Calibration type, the following menu items
are available:

KNOWN POINT 1/5

”: point 1 of 5, usually 0% of the transducer’s full-scale capac-

ity.

KNOWN POINT 2/5

”:

point 2 of 5, usually 25% of the transducer’s full-scale capac-

ity.

KNOWN POINT 3/5

”:

point 3 of 5, usually 50% of the transducer’s full-scale capac-

ity.

KNOWN POINT 4/5

”:

point 4 of 5, usually 75% of the transducer’s full-scale capac-

ity.

KNOWN POINT 5/5

”:

point 5 of 5, usually 100% of the transducer’s full-scale

capacity.

These menu items are only available with the Known-Load Calibration types.

NOTICE

To insure both correct operation of the transducer and application of the load,
the instrument expects the voltage or current applied at each known-load point
to be increasing. For example, the load applied at Known-Load Point 2/2 must
cause the transducer to produce a more positive voltage or current than at
Known-Load Point 1/2.