Honeywell SC1000 User Manual

Page 118

Advertising
background image

page 118 008-0608-00

If the

CALIBRATION TYPE

is 5-Point Known Load Calibration...

The display will read

DOING 5POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/5

and

DISPLAY.

UNITS

). When you have applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00250 UNIT

(where 00250 and UNIT are

previously entered as

KNOWN POINT 2/5

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are

previously entered as

KNOWN POINT 3/5

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00750 UNIT

(where 00750 and UNIT are

previously entered as

KNOWN POINT 4/5

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 5/5

and

DISPLAY. UNITS

). When you have

applied this load to the transducer, press [ENTER].

The display will now read

RETURN PNTS.=NO

. If you press [ENTER], the display

will indicate

DONE

and the instrument will return to the RUN mode. If you use

the [UP] button and [ENTER] to select

RETURN POINTS=YES

, you may now apply

the same loads again but in reverse order. This will allow the instrument to
measure and partially eliminate any hysteresis that may be present in the
transducer. After the loads have been applied again, the display will indicate

DONE

and the instrument will return to the RUN mode.

Advertising
This manual is related to the following products: