page 72 008-0608-00

applied this load to the transducer, press [ENTER].

Upon completion, the display will indicate

DONE

and the instrument will return to the

RUN mode.

If the

CALIBRATION TYPE

is 5-Point Known Load Calibration...

The display will read

DOING 5POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where

00000 and UNIT are previously entered as

KNOWN POINT 1/5

and

DISPLAY. UNITS

).

When you have applied this load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00250 UNIT

(where 00250 and UNIT are pre-

viously entered as

KNOWN POINT 2/5

and

DISPLAY. UNITS

). When you have applied this

load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are pre-

viously entered as

KNOWN POINT 3/5

and

DISPLAY. UNITS

). When you have applied this

load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00750 UNIT

(where 00750 and UNIT are pre-

viously entered as

KNOWN POINT 4/5

and

DISPLAY. UNITS

). When you have applied this

load to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are pre-

viously entered as

KNOWN POINT 5/5

and

DISPLAY. UNITS

). When you have applied this

load to the transducer, press [ENTER].

The display will now read

RETURN PNTS.=NO

. If you press [ENTER], the display will

indicate

DONE

and the instrument will return to the RUN mode. If you use the [UP]

button and [ENTER] to select

RETURN POINTS=YES

, you may now apply the same

loads again but in reverse order. This will allow the instrument to measure and
partially eliminate any hysteresis that may be present in the transducer. After the
loads have been applied again, the display will indicate

DONE

and the instrument

will return to the RUN mode.