4 calibration procedure – Honeywell SC1000 User Manual

Page 107

Advertising
background image

SC Series Instruction Manual page 107

Hig

h-L

ev

el

In

pu

t C

han

nel

12

12.4 Calibration Procedure

If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23. A listing of all menu items is given in
“Setup Menu Reference” on page 157.

Step 1: Wire the transducer to the channel’s connector.

See “Wiring” on page 98 for details.

Step 2: Set the Excitation and Signal jumpers appropriate for the transducer,

amplifier or DC-DC LVDT.
See “Wiring” on page 98 and “Excitation and Signal Jumpers” on page 106.

Step 3: Enter the

CALIBRATION TYPE

.

There are two methods that can be used to calibrate the transducer to the
Input Channel. Each has advantages and disadvantages as described in
“CALIBRATION TYPE Menu Item” on page 113. It is important to know your
application in order to select the appropriate calibration type.

Step 4: Enter the

CALIBRATION DATA.

Otherwise, consult the Certificate of Calibration for the transducer when
entering information in the

CALIBRATION DATA

sub-menu.

Step 5: Perform the calibration.

Using the

CALIBRATE

menu item starts the calibration process. You will be

prompted to apply loads to the transducer as required.