used to compensate for the non-linearity in the transducer. The usual loads
used are zero scale, half scale and full scale, but you are not restricted to
these loads.

TYPE= 5 POINT CAL

” means 5-Point Known Load Calibration. You are prompted

to apply the loads to the transducer that were entered in the “

KNOWN POINT 1/5

”,

KNOWN POINT 2/5

”, “

KNOWN POINT 3/5

”, “

KNOWN POINT 4/5

” and “

KNOWN POINT 5/5

” regis-

ters. This technique can be used to compensate for the non-linearity in the
transducer. The usual loads used are zero scale, one-quarter scale, half
scale, three-quarters scale and full scale, but you are not restricted to these
loads.

The selection of the calibration type affects what menu items appear in the

CALI-

BRATION DATA

sub-menu.