Error 29,

S.EE VERIFY FAIL

A verification failure occurred during a write operation to the channel’s EEPROM.
Restart the instrument.

Error 30,

S.EE ERASE NAK

The channel’s EEPROM could not be detected during an erase operation. Check
that the channel’s address jumpers are set properly.

Error 31,

BLANK IDENT

The channel does not appear to have been installed properly. This occurs if you
swap Input or Output boards without using

SYSTEM MENU

->

INSTALL CHANNEL

to format

the channel’s EEPROM.

Error 32,

BOGUS EE ADDR

An invalid address has been passed to the programmer. Restart the instrument. If
this problem persists, contact our Service department with the information found
in the

SYSTEM MENU

->

SOFTWARE REV.

menu item.

Error 33,

SIG.MOD WIRES

The instrument has detected that the (+)Signature and (-)Signature pins (labeled
as “+MEM” and “-MEM) are likely reversed.

Error 34,

SIG.MOD MISMATCH

The excitation type byte found inside a Signature Module indicates that the chan-
nel is not designed to operate with that type of transducer.

Error 35,

BOGUS MENU TYPE

A menu contains an invalid menu type. Contact Sensotec with the information
found in the

SYSTEM MENU

->

SOFTWARE REV.

menu item.

Error 36,

BAD EE TABLE

The instrument doesn’t understand the EEPROM type setting for the channel.
Contact Sensotec with the information found in the

SYSTEM MENU

->

SOFTWARE REV.

menu item.

Error 37,

BAD PERSONALITY

The channel doesn’t understand the given card type. Contact Sensotec with the
information found in the

SYSTEM MENU

->

SOFTWARE REV.

menu item.

Error 38,

BAD FORMAT

The channel’s EEPROM memory contains a display setting which is not under-
stood by the instrument. Check that every item in the channel’s

DISPLAY SETUP

menu has a valid setting.

Error 39,

WRONG HARDWARE

The channel type expected by the instrument for the channel number is not the
channel type that is physically installed. If you have changed the address jumpers
on one or more channels, replace them in their original configuration.

Error 41,

SHUNT CAL ERROR

The application of the shunt resistor did not affect the transducer’s output. Check
the wiring and the transducer output with a voltmeter to verify proper operation.

If this is a Strain Gage Channel, make certain that the shunt calibration resistor
installed on the circuit board is the correct value per the transducer’s Calibration
Record.

If this is a High-Level Input Channel, make certain that the Shunt Cal 1 and Shunt
Cal 2 pins are connected to the correct pins of the transducer/amplifier.

Error 43,

BAD IDENT CRC

A checksum failure has occurred when reading a channel’s EEPROM. Restart
the instrument.

Error 44,

BAD FUNCTION

The instrument has requested that the channel perform a function that it is not
capable of. Restart the instrument.

Error 45,

A/D NAK

Hardware error. The channel’s Analog-to-Digital converter is not communicating
with the channel’s microprocessor. Restart the instrument.

Error 46,

SENSE WIRES

The Strain Gage Channel’s (+)Sense and (-)Sense pins (labeled as “+SEN” and “-

Advertising
This manual is related to the following products: