SC Series Instruction Manual page 65

Strain-

Gage

Input

C

ha

nnel

10

10.5.2

DISPLAY SETUP

Sub-Menu

This menu controls how data values are displayed by the channel and transmitted
via serial communications.

DISPLAY. DIGITS

Menu Item

Selects the number of digits displayed by the channel. The choices are:

5 DIGIT-BIPOLAR

” displays both positive and negative numbers with five full dig-

its (±99999).

6 DIGIT-UNIPLOAR

” displays positive numbers with six full digits (999999) and

negative numbers with five full digits (-99999).

7 DIGIT UNIPOLAR”

displays positive number with seven full digits (9999999)

and negative numbers with six digits (-999999).

This menu item may be automatically updated by a transducer’s Signature Mod-
ule
.

DISPLAY. DECPT

Menu Item

Selects the decimal point location on the channel’s display and serial communica-
tions output. Use the [UP] and [DOWN] buttons to move the decimal point to the
right and left.

This menu item may be automatically updated by a transducer’s Signature Mod-
ule
.

DISPLAY. COUNT-BY

Menu Item

Determines by what value the display increments or decrements. This will make
the display less sensitive to a noisy signal. The choices are:

00001

00002

00005

00010

00020

00050

00100

00200

This menu item may be automatically updated by a transducer’s Signature Mod-
ule
.

DISPLAY. UNITS

Menu Item

Specifies the four character label that is displayed to the right of the channel’s val-
ues.

When a character position is flashing press the [UP] or [DOWN] button to change
the character. Press [ENTER] to advance to the next character.

This menu item may be automatically updated by a transducer’s Signature Mod-
ule
.

DISPLAY. AVERAGE

Menu Item

Controls the speed with which the channel’s display values will update. Display
averaging does not affect the channel’s analog output or its peak/valley detection,
which will proceed at the speed selected by the ”

OPERATION

->

FREQ. RESPONSE

” menu

item. The choices are:

ON

” means that the display will update four times each second. The channel’s

values will be averaged for 1/4 second, then displayed.

OFF

” means that the channel’s display will update as quickly as possible.

NOTICE

This menu item doesn’t change the mathematical scaling of the channel’s val-
ues; that can be changed by altering the “

CALIBRATION DATA

->

FULL SCALE VALUE

menu item and then performing a re-calibration.