Strain-

Gage

Input

C

ha

nnel

10

10.5.4

CALIBRATION TYPE

Menu Item

This chooses the type of calibration technique to be used. There are three meth-
ods that can be used to calibrate the transducer to the Input Channel. Each has
advantages and disadvantages as described in the table below.

It is important to know your application in order to select the appropriate cali-
bration type. mV/V (“millivolt-per-volt”) Calibration is recommended for most
applications because it doesn’t require known loads and it allows easy swapping
of transducers equipped with Signature Calibration. When the best possible
accuracy is required, Known-Load Calibration is recommended.

The choices for this menu item are:

TYPE= SHUNT CAL

” means Shunt Calibration. First, you are prompted to apply

the load entered in the “

ZERO SCALE VALUE

” register. Next, the instrument inserts

Table 4-4: Comparison of Calibration Types

Shunt

Calibration

mV/V

Calibration

2-, 3-, or 5-

point

Known Load

Calibration

RECOMMENDED...

...for most applications

no

yes

no

...when frequently swapping

transducers

yes

yes

no

...when best possible accuracy

required

no

no

yes

PROS and CONS

Relative accuracy

good

better

best

Requires actual, calibrated loads

to be applied

no

no

yes

Automatically calibrates with Sig-

nature Calibration transducers

yes

yes

no

Max. distance of transducer from

instrument w/out sense wires

15’ [5m]

50’ [15m]

300’ [100m]

Max. distance of transducer from

instrument with sense wires

15’ [5m]

100’ [30m]

300’ [100m]

CALIBRATION TECHNIQUE

Zero-scale point from...

applied load

applied load

applied load

Full-scale point from... transducer

output when

shunt resistor

applied

internal refer-

ence

applied load

Linearity correction from...

data stored in

Signature

Module (if

available)

data stored in

Signature

Module (if

available)

2-point: none

3- or 5-point:

applied load