11.2 Wiring

Connect your transducer to an AC/AC-LVDT Input channel by wiring it to the 12-pin
connector of that channel. The Customer Information Sheet that shipped with the
instrument describes which cards are installed in each channel. The pin-out for this
connector is shown on the following table.

The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument.

Figure 11-1: Connection of Four- or Five-wire AC/AC-LVDT

Table 5-5: AC/AC-LVDT Input Channel Pin Connections

Pin

Label

Function

Reference Pin

1 (top)

+EXC

(+)Excitation

10

2

N/C

No connection

3

N/C

No connection

4

-EXC

(-)Excitation

10

5

+SIG

(+)Signal

10

6

-SIG

(-)Signal

10

7

+OUT

Analog Output

8

8

-OUT

Analog Return

-

9

N/C

No connection

10

DGND

Digital Ground

-

11

AUX1

Auxiliary Function 1

(connect to pin 10 to activate)

10

12

(bottom)

AUX2

Auxiliary Function 2

(connect to pin 10 to activate)

10