6 calibrate menu item, Menu item, Notice – Honeywell SC1000 User Manual

Page 89

Advertising
background image

SC Series Instruction Manual page 89

AC

/A

C-L

VDT I
nput Ch

annel

11

11.5.6

CALIBRATE

Menu

Item

This menu item performs a calibration according to what was entered in the

CALI-

BRATION TYPE

and

CALIBRATION DATA

menu items.

If the

CALIBRATION TYPE

is 2-Point Known Displacement Calibration...

The display will read

DOING 2POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/2

and

DISPLAY.

UNITS

). When you have applied this displacement to the transducer, press

[ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 2/2

and

DISPLAY. UNITS

). When you have

applied this displacement to the transducer, press [ENTER].

Upon completion, the display will indicate

DONE

and the instrument will return to

the RUN mode.

If the

CALIBRATION TYPE

is 3-Point Known Displacement Calibration...

The display will read

DOING 3POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/3

and

DISPLAY.

UNITS

). When you have applied this displacement to the transducer, press

[ENTER].

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are

previously entered as

KNOWN POINT 2/3

and

DISPLAY. UNITS

). When you have

applied this displacement to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 01000 UNIT

(where 01000 and UNIT are

previously entered as

KNOWN POINT 2/3

and

DISPLAY. UNITS

). When you have

applied this displacement to the transducer, press [ENTER].

Upon completion, the display will indicate

DONE

and the instrument will return to

the RUN mode.

If the

CALIBRATION TYPE

is 5-Point Known Displacement Calibration...

The display will read

DOING 5POINT CAL

, and prompt you to

APPLY 00000 UNIT

(where 00000 and UNIT are previously entered as

KNOWN POINT 1/5

and

DISPLAY.

UNITS

). When you have applied this displacement to the transducer, press

[ENTER].

The display will read

WORKING

, then

APPLY 00250 UNIT

(where 00250 and UNIT are

previously entered as

KNOWN POINT 2/5

and

DISPLAY. UNITS

). When you have

applied this displacement to the transducer, press [ENTER].

The display will read

WORKING

, then

APPLY 00500 UNIT

(where 00500 and UNIT are

NOTICE

Before performing a calibration,
the transducer must be connected to the instrument, the

CALIBRATION TYPE

must

be selected (see “CALIBRATION TYPE Menu Item” on page 87) and the

CALIBRA-

TION DATA

must be entered (see “CALIBRATION DATA Sub-Menu” on page 88).

NOTICE

For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.