HP RM500SL User Manual

Page 34

Advertising
background image

0611

RM500SL User’s Guide Version 2.8

Page 34

7.9

Calibration of HIT reference microphone

FastFacts 7.9: Calibration of HIT reference microphone

Remove any coupler from the coupler microphone before calibration.

1) Position the HIT reference microphone opening concentric with the coupler

microphone opening and about 1 – 2 mm from as shown below.

2) Close and seal the test chamber by tightening the latch at the front of the lid.

3) Press <Tests>, then highlight and <PICK> [Calibrate] from the Hearing

Instruments Test column.

4) Highlight, then <PICK> [Daily] or [Weekly] to select the calibration interval.

5) Highlight and <PICK> [Calibrate].

See 7.8: HIT calibration facts

It is recommended that you print a copy of the initial calibration curve and compare
it with your daily or weekly curve. Any change in this curve indicates a drift in
either the coupler or the reference microphone which needs further investigation.
See

7.10: Calibration check for coupler microphone

for an additional check.