16 Insertion Gain

16.1 Section overview

FastFacts 16.1: Insertion gain

16.2 Insertion gain in SPL

FastFacts 16.2: Insertion gain in SPL

Insertion gain is the difference between aided and unaided ear canal SPL (REIG =
REAG - REUG). It inherently assumes a sound-field audiogram. For non-linear
hearing aids, REIG is stimulus-dependent and the dependency is specific to each
hearing aid. Insertion gain fitting methods ignore these facts. Speechmap is
recommended for fitting all non-linear hearing instruments.

Refer to

10: Real-Ear Measurement Setup

and

11: REM Screen Setup

for

terminology and information on preparing the system for use and for proper
positioning of the client and the probe tube.

This section covers the following topics:

16.2: Insertion gain in SPL

16.6: REAR measurement procedure

16.3: Insertion gain in HL

16.7:SII calculation in Insertion gain

16.4: Audiometric data entry

16.5: REUR measurement procedure

See

10: Real-Ear Measurement Setup

and

11: REM Screen Setup

1) Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list.

2) Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body.

This changes the CORFIG used for 2cc targets.

3) Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R,

POGO II, Berger, Libby, User, None