RM500SL User’s Guide Version 2.8

Page 67

16.5 REUR measurement procedure

FastFacts 16.5: REUR measurement procedure

Performing a REUR test erases all existing REAR and REIG data for the current
test ear. REUR should be measured only if Transducer = Soundfield.
Otherwise it is more accurate to use average. See

16.4: Audiometric data entry

.

1) Press <Tests>, then highlight and <PICK> [Insertion gain] from the REM list.

2) Referring to

10: Real-Ear Measurement Setup

and

11: REM Screen Setup

,

prepare the system, position the patient and insert the probe tube.

3) If REUR is set to Average, highlight and <PICK> [Audiometry] then highlight

and <PICK> [REUR] to toggle to [Measure]. Press <Continue> to advance to
the Threshold entry poster. You may enter threshold data at this time (see

16.4:

Audiometric data entry

) or press <Continue> to proceed with the measurement.

4) Highlight and <PICK> REUR (below REAR/REIG test 3).

5) Press <Continue> to accept the default 55 dB SPL Pink noise stimulus and

capture the REUR curve OR highlight and <PICK> [Stimulus] and [Level] to
select other values. These choices have no impact on the resulting REIG.

6) If there is a notch in the 4000 to 8000 Hz region, try advancing the probe tube 2

mm and repeating step 3. If the notch moves to a higher frequency, repeat this
step until the notch moves beyond the highest frequency of interest.

Because it is analyzed in 1/3 octave
bands, the Noise stimulus produces a
REUR 12 dB below that produced by the
Swept stimulus at the same overall level.

The Average REUR used is shown at
right. Measured REUR will vary from this.

For small children the peak may shift to the right and become sharper. Drainage
tubes and perforated ear drums will cause large differences - usually around 1 kHz.