HP RM500SL User Manual

Page 61

Advertising
background image

0611

RM500SL User’s Guide Version 2.8

Page 61

15.3 RECD coupler response

FastFacts 15.3: RECD coupler response

Prepare the system for HIT tests. See

2.3: Microphone connection

,

7.9:

Calibration of HIT reference microphone

. Connect the optional RECD transducer

to the RECD jack to the left of the test chamber.

1) Press <Tests> then highlight and

<PICK> [RECD] from the REM
tests list.

2) If a valid coupler response exists,

go to

15.4:RECD real-ear

response.

3) If no coupler response exists or the

measurement interval has been
exceeded, connect the RECD
transducer to the BTE coupler.

4) Highlight and <PICK> [Measure coupler] and follow the setup instructions.

5) Press <Continue> to generate the coupler response curve (green). When this

curve is stable, press <Continue> to save it.

6) Highlight and <PICK> [Daily] or [Weekly] to toggle the measurement interval for

the coupler response.

For comparison purposes, an age-related average RECD is shown as a dotted
curve on the screen. If you have accessed RECD from the REM tests list, you may
highlight and <PICK> the Age box to change the age for this reference curve.