Manual Supplement

4/08 1

Change #1,

46470, 46697, 47714, 48007, 49680

Replace pages 31 through 35, with the following:

Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes

................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic

Note

AC Pulsed Biphasic waveform has not been approved in the United States.

Autoranged Measurement

.................................... 0.1 to 600 J

Accuracy

0.1 to 360 J .....................................................

±(1 % of reading + 0.1 J)

360 to 600 J ....................................................

±(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is

±(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

Resistance .......................................................... 50

Ω

Accuracy .........................................................

±1 %, non-inductive (<2 μH)

Pulse trigger level

................................................. 20 V

Pulse width

Range ................................................................. 1.0 to 50.0 ms