Impulse 6000D/7000DP Users

Manual Supplement

4/08 3

Scope Output

Autorange ........................................................... 2000:1, 400:1 and 80:1: dependant on the range

Waveform Playback

Output ............................................................. BNC
Output impedance........................................... 50

Ω

Amplitude Accuracy ........................................

±5 %

Charge Time Measurement

Range ................................................................. 0.1 to 100.0 s
Accuracy .............................................................

±0.05 s, typical

Synchronization Test (Elective Cardioversion)

Delay Time Measurement

Timing window ................................................ ECG R-wave peak to the defib pulse peak
Range ............................................................. -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380

ms following the R-wave peak.

Resolution ....................................................... 1 ms
Accuracy .........................................................

±1 ms

ECG waves

Normal Sinus Rhythm (NSR) .......................... 10 to 180 (by 1) BPM
Atrial fibrillation................................................ Coarse and fine
Monomorphic Ventricular Tachycardia ........... 120 to 240 (by 5) BPM
Asystole .......................................................... Flat line

Automated Defibrillator Test ECG Waves

Normal Sinus ...................................................... 10 to 300 (by 1) BPM