Manual Supplement

Impulse 6000D/7000DP Users

4

3/09

Ventricular Fibrillation ......................................... Coarse and fine
Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia ................. 5 types
Asystole .............................................................. Flat line

ECG Waves

ECG General

Lead configuration .............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ..................................... 1000 Ω (nominal)
Rate accuracy.....................................................

±1 % of nominal

ECG Amplitudes

Reference Lead...................................................Selectable, Lead II (default) or Lead I
Settings................................................................0.05 to 0.45 (by 0.05) mV

0.5 to 5.0 (by 0.5) mV

Accuracy (All Performance waves and Normal Sinus R waves)

Lead II..................................................................±2 %
All other leads......................................................±5 %
Defib paddles.......................................................±5 %

Amplitude of ECG signals relative to amplitude setting (in percent)

Lead II reference

Performance waves and R wave detection

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

70 %

100 %

30 %

100 %

100 %

100 %

100 %

100 %

100 %