Impulse 6000D/7000DP Users

Manual Supplement

8/08 5


Normal Sinus waves

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

70 %

100 %

30 %

24 %

48 %

100 %

120 %

112 %

80 %

Lead I reference

Performance waves and R wave detection

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

100 %

150 %

50 %

100 %

100 %

100 %

100 %

100 %

100 %

Normal Sinus waves

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

100 %

150 %

50 %

24 %

48 %

100 %

120 %

112 %

80 %

ECG Normal Sinus

Rates................................................................... 10 to 360 (by 1) BPM

ECG High Level Output (BNC Jack)

Amplitude

Range ............................................................. 0.5 V per mV of reference lead setting
Accuracy .........................................................

±5 %

Output Impedance .............................................. 50 Ω

ECG on Defibrillator Input Load

Same as the LEAD II amplitude but limited to ±4 mV

ECG Performance Waves

Square wave ....................................................... 2.0 and 0.125 Hz
Triangular wave .................................................. 2.0 and 2.5 Hz