Manual Supplement

Impulse 6000D/7000DP Users

2

8/08

Accuracy.............................................................

±0.1 ms

Voltage

Range ................................................................. 20 to 5000 V
Accuracy.............................................................

±(1 % of reading + 2 V)

Current

Range ................................................................. 0.4 to 100.0 A
Accuracy.............................................................

±(1 % of reading + 0.1 A)

Tilt (biphasic and pulsed biphasic)

Range ................................................................. 1 % to 99 %

Accuracy.............................................................

±1 digit

Interphase delay (biphasic and pulsed biphasic)

Range ................................................................. 0.1 ms to 9.9 ms

Accuracy.............................................................

±0.1 ms

Frequency (pulsed biphasic only)

Range ................................................................. 2000 Hz to 8000 Hz

Accuracy.............................................................

±1 % of reading

Duty cycle (pulse biphasic only)

Range ................................................................. 1 % to 99 %

Accuracy.............................................................

±1 digit

Sample rate

........................................................... 250 kHz (4

μs sample)

Maximum Average Power

.................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes