Impulse 6000D/7000DP Users

Manual Supplement

4/08 7

Accuracy ......................................................... ±(10 % of setting + 0.2 mV)

Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)

Lead II reference

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

67 %

100 %

33 %

67 %

67 %

67 %

67 %

67 %

67 %

Lead I reference

Lead

# I II III V1 V2 V3 V4 V5 V6

Ref. Amp.

100 %

150 %

50 %

100 %

100 %

100 %

100 %

100 %

100 %

Arrhythmia Selections

Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)

Demand .......................................................... 30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA

Supraventricular

Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia