POSSIBLE CAUSE

Concentrate valve plugged

Concentrate outlet line

restricted

Flow meter inaccurate

Dirty concentrate valve

Recycle or concentrate valve

plugged

Inaccurate pressure gauge

Restricted or reduced

permeate flow rate

Restricted flow after

pump outlet

PROBLEM

Low concentrate flow

rate, normal or higher

than normal pressure

High operating pressure

Excessive pressure drop

[over 50 psig (3.0 barg)]

(high primary pressure -

low final pressure)

SOLUTION

Remove the concentrate valve step

and/or disassemble the plumbing.

Clean the valve

Examine the concentrate line for

obstructions or kinks, repair or

replace the tubing.

Check the flow rate manually with a

stop watch and calibrated

container.

Disassemble and clean the

plumbing to the valve

Disassemble the plumbing to the

recycle valve and remove foreign

particles. Check the concentrate

valve stem.

Replace or calibrate the gauge as

required.

See possible causes for low

permeate rate.

Check for blockage of the concen-

trate flow at the inlets and outlets of

the membrane element housings.

T R O U B L E S H O O T I N G G U I D E