Spring adapter for the water
connection

The spring adapter for the water
connection must be properly engaged
when a basket or the injector wagon is
inserted into the machine. It must be
4 - 5 mm higher than the water inlet in
the machine.

If this is not the case, the adapter
needs to be adjusted as follows:

^ Loosen the locking ring a.
^ Move the adapter up to the proper

height b.

^ Tighten the locking ring c.

Adjusting the upper basket

The upper basket can be adjusted
above and below the middle position,
by 2 cm.

Depending on the height selected and
which inserts are used, labware of
various heights can be arranged in the
baskets.

To adjust the upper basket:

^ Pull the upper basket out as far as it

will go, lift it up off the runners and
remove.

^ Use a wrench to unscrew the roller

bearings on either side of the basket,
then reposition as required.

Loading the machine