Cleaning the water inlet filters

To protect the water inlet faucet, filters
are incorporated in the intake hose
attached to the water supply. If the
filters are soiled they must be cleaned
to ensure sufficient water intake to the
wash cabinet.

,

The plastic housing of the water

connection contains electrical
components and should not be
submerged or run through water.

Cleaning the filters

^ Disconnect the machine from

electricity (unplug it or "trip" the
circuit breaker).

^ Close the water supply and unscrew

the water inlet faucet.

^ Carefully remove the rubber seal.
^ Take out the filters with needle nose

pliers.

^ Rinse the large area filter 1 and fine

filter 2 under running water, and
replace if necessary.

^ Return filters and seal. Make sure

they are seated correctly.

^ Reconnect the hose to the water

faucet, checking that it is not
cross-threaded.

^ Open the water supply.
^ Check for leaks.