Cleaning the drain pump and
non-return valve

If water has not been pumped away at
the end of a program the drain pump or
the non-return valve may be blocked.

^ Turn off the machine.
^ Remove the filter combination out of

the wash cabinet.

^ Tilt the locking clamp to the side.
^ Pull out the non-return valve and

rinse well under running water.

,

Use proper eye protection when

removing the non-return valve.

The drain pump is beneath the
non-return valve (see arrow).

^ Before refitting the non-return valve,

check that the drain pump is not
blocked. Spin the propeller several
times in both directions to check for
obstructions.

^ Carefully refit the non-return valve

and secure it with the locking clamp.
Refit and lock the filter in place.