Cleaning the spray arms

The spray arms can get clogged.
Check daily and clean if necessary.

Removing the spray arms

^ Remove the baskets.

Spray arm at the upper basket or
mobile unit (if available):

^ Loosen the knurled nut and remove

the spray arm.

Metal knurled nuts have a left-hand
thread. Ceramic knurled nuts have a
right-hand thread.

^ Unscrew the upper spray arm.
^ Loosen the knurled nut of the lower

spray arm and remove.

^ Use a sharp pointed object to push

particles into the spray arm jets and
rinse thoroughly under running water.

^ Refit the spray arms in reverse order

after cleaning.

After replacing the spray arms
check that they can rotate freely.