Environmental conditions – Sanyo MDF-593C User Manual

Page 10

Advertising
background image

9

ENVIRONMENTAL CONDITIONS

This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1):


1. Indoor use;

2.
Altitude up to 2000 m;

3.
Ambient temperature 5

o

C to 40

o

C


4.
Maximum relative humidity 80% for temperature up to 31

o

C decreasing linearly to 50% relative

humidity at 40

o

C;


5.
Mains supply voltage fluctuations not to exceed

10% of the nominal voltage;


6.
Other supply voltage fluctuations as stated by the manufacturer;

7.
Transient over-voltages according to Installation Categories (Over-voltage Categories) II; For mains
supply the minimum and normal category is II;

8.
Pollution degree 2 in accordance with IEC 664.

Advertising