OPERATING INSTRUCTIONS

Back-up system

WARNING

As with any equipment that uses CO

2

gas, there is a likelihood of oxygen depletion in the vicinity

of the equipment. It is important that you assess the work site to endure there is suitable and

sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a

safe environment must be considered. These may include atmosphere monitoring and warning

devices.

The freezer is provided with an automatic liquid CO

2

gas injection device as a back-up system. This

freezer prevents the chamber temperature from going up by injecting the liquid CO

2

gas when the power

supply is disconnected (power failure, disconnection of power cord, breaker OFF) or in the case of failure

of freezer itself. The liquid CO

2

gas is injected with the activation of solenoid valve energized by battery

when the chamber temperature reaches the alarm temperature.

Following shows the procedure for setting the back-up system.

1. Setting of liquid CO

2

gas cylinder

By using the joint and pipe enclosed with the

freezer, connect the liquid CO

2

gas cylinder to the

joint of the back-up system. For this setting,

consult with a qualified gas supplier or Sanyo sales

agency.

2. After setting the liquid CO

2

gas cylinder, operate

the freezer until the chamber temperature reaches

the required level.

3. Set the warning indicator of the recorder at the

temperature higher than -70

o

C. The back-up

system is operated continuously if the warning

indicator is set at the temperature lower than -70

o

C.

This means the liquid CO

2

gas is consumed very

quickly.

4. Switch on the back-up system.

5. Make sure that liquid CO

2

gas spouts into the freezer chamber by pressing the back-up test switch.

Note:

• The liquid CO

2

gas cylinder loses its cooling capacity at speed when the ambient temperature is over

31

o

C. Install the liquid CO

2

gas cylinder in the cool environment. And the duration of back-up time per

one liquid CO

2

gas cylinder varies depending on the ambient temperature. Refer to “Installation of

back-up system” enclosed with the freezer for the available back-up time.

• The liquid CO

2

gas cylinder should be a siphon type.

• Use the pipe encloses with the freezer for the setting. (The extension of the pipe is not permitted

because of cooling capacity.)

Battery

Pipe

Liquid CO2
gas cylinder

Solenoid valve