TEMPERATURE RECORDER

8. Remove the sensor cover at the rear of the chamber wall. (Fig.

4)

9. Pass through the hole under the temperature control sensor. (Fig.

5) Tool for the hole: Screw driver (example) Hole depth:

135mm (5.3inch) Hole diameter: 13mm (0.5inch)

10. Route the recorder sensor into the chamber by sassing the hole.

(Fig. 4)

11. Remove the sensor support and position the recorder sensor and

then replace the stopper. (Fig. 4)

12. Seal around the hole for recorder sensor completely by applying

the sealing compound (RTV) to outside and inside of the chamber.

(Fig. 4)

13. Connect the power cord of the recorder to the terminal as shown

in Fig. 6.

N

Neutral

L

Live

E

Earth

14. Fix the grille with a screw and replace the side cover.