Operating instructions, Temperature setting, Remote alarm terminal – Sanyo MDF-593C User Manual

Page 17

Advertising
background image

16

OPERATING INSTRUCTIONS

Temperature setting

Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the

table. The example in the table is based on the assumption that the temperature is -70

o

C.

Note: The unit is set at the factory that the chamber temperature is -85C.

Table 1. Basic operation sequence (Example: Chamber temperature -70

o

C)

Description of operation

Key operated

Indication after operation

1 Turn on the power switch.

-----

The current chamber temperature is

displayed.

1

Press mode change key when

changing the set temperature

during operation.

PV/SV

The second digits blinks.

2

Set the second digit to 7 with the

numerical value shift key.

When pressed, the figure of settable

digit changes.

3

By pressing the digit shift key, set

the first digit.

The first digit blinks.

4

Set the first digit to 0 with the

numerical value shift key.

When pressed, the figure of settable

digit changes.

5 Press enter key.

ENT

Set temperature is memorized and

the current chamber temperature

is displayed.

Note: The temperature set mode returns to the temperature display mode automatically when 90

seconds has passed without any key operation.

Remote alarm terminal

The terminal of the remote alarm is located at rear of the machinery room, upper side of the electric box.

The signal is contact output. The recommended contact capacity is between 2 A (DC 30V) .

Contact output: : normal close connect with N.O. and COM.

Consult our dealer or our office for installation of the remote alarm.