Operating instructions, Alarm and safety functions – Sanyo MDF-593C User Manual

Page 20

Advertising
background image

19

OPERATING INSTRUCTIONS

Alarm and safety functions

This unit has the alarm and safety functions shown in table below, and also self diagnostic function.

Table 4. Alarm and safety function

Alarm & safety

Situation

Indication

Buzzer

Safety operation

High temperature
alarm

When the chamber temp. is
approx. 10

o

C (or 15

o

C) higher

than the set temp.

Alarm lamp blinks.

Intermittent
tone with
12
min. delay

Remote alarm
with 12 min. delay

Power failure
alarm

At power failure
When power cord is
disconnected.

Alarm lamp blinks.

Intermittent
tone

Remote alarm

Filter check

When the condenser filter
is clogged.

Filter check lamp lights.

-----

-----

Auto-return

When there is no key
pressing in each setting
mode for 90 seconds.

Chamber temp. is
displayed.

-----

Finishing of each
setting mode

If the thermal sensor is
disconnected.

Alarm lamp blinks.
E01 and 50

o

C is

displayed alternately.

If the thermal sensor is
shirt-circuited.

Alarm lamp blinks.
E01 and -170

o

C is

displayed alternately.

Intermittent
tone

Remote alarm
Continuous
operation

If the cascade sensor is
disconnected.

Alarm lamp blinks.
E03 and chamber temp.
is displayed alternately.

If the cascade sensor is
shirt-circuited.

Alarm lamp blinks.
E04 and chamber temp.
is displayed alternately.

Intermittent
tone

Remote alarm

If the filter sensor is
disconnected.

Alarm lamp blinks.
E05 and chamber temp.
is displayed alternately.

Sensor
abnormality

If the filter sensor is
short-circuited.

Alarm lamp blinks.
E06 and chamber temp.
is displayed alternately.

Intermittent
tone

Remote alarm

Note

The filter check lamp sometimes lights at the start-up under high ambient temperature. The lamp will is

off when the chamber temperature is getting lower.

Operation after power failure

The set value (setting of chamber temperature and alarm temperature) is memorized by nonvolatile

memory. Accordingly, the freezer resumes the operation with setting before power failure.