VOR/ILS Analyzer EVS200

0796.1800.02

E-8

10

Chapter 2: Preparation for Operation

External AF input

Via the AF input (AF-EXT) at the rear side the unit can be fed with an AF signal for further analysis of
typical AF parameters (e.g. level, frequency). This is very applicable for receivers which cannot per-
form AF analysis.

input level:

approx.: 1 to 2 V

rms

/ 50

e.g. VOR/ILS Analyzer

Þ=

external receiver

Connection AF-EXT of the VOR/ILS Analyzer EVS200 with
the AF output of an external receiver

DSP output

On the DSP-OUT at the rearside e.g. a XY-tracer can be connected. The scaling of the XY values
can be set for any mode (localizer/glideslope) in the setup (DDM Y / t-RANGE) or in ILS-Mode
(Y / t-SETUP).

output signals in VOR- and ILS-Mode:

ILS-Mode

VOR-Mode

Display 1

normalized DDM-values
(DC-voltage)
Range 1 to 4

no output

Display 2

voice-frequency
(300 to 3000 Hz)

voice-frequency
(300 to 3000 Hz)

scaling in the setup:

Localizer

Glideslope

Range 1

Range 2

Range 3

Range 4

0.0

±

25%

0

±

0.25 DDM

0.0

±

2.5%

0

±

0.025 DDM

0.0

±

2.58%

0.0

±

0.0258 DDM

0.0

±

50%

0.0

±

0.5 DDM

0.0

±

50%

0.0

±

0.5 DDM

0.0

±

5%

0.0

±

0.05 DDM

8.75

±

5%

0.0875

±

0.05 DDM

17.5 +5%

0.175 +0.05 DDM

(corresponding

DC voltage)

(0 V

±±±±

5 mV)

0 DDM

(1 V

±±±±

5 mV)