Procom VOR/ILS Analyzer EVS200 User Manual
Page 55

VOR/ILS Analyzer EVS200
0796.1800.02
E-8
46
Chapter 3: Operation
Command D2
command
function
D2
<Display-information>
switching over to Display2
The command D2 enables the switching over to the content of display2 in the VOR / ILS mode.
This is the prediction for reading out the data of the corresponding mode (VOR/ILS).
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
input:
D2 <CR>
d2 <CR>
READY <CR/LF>
READY <CR/LF>
Command AF8
command
function
AF8
<AF-frequency ID>
AF-frequency ID query
The command AM8 reads the actual AF frequency (ID) in the ILS / VOR mode.
Prediction for reading out the data of the corresponding mode (VOR/ILS) is the
switching over to display2.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
AF8 <CR>
af8 <CR>
e.g.
1020.0Hz <CR/LF>
e.g.
1020.0Hz <CR/LF>
Command AM8
command
function
AM8
<AM-modulation depth (ID)>
query of AM-modulation depth (ID)
The command AM8 reads the actual AM-modulation depth (ID) in the ILS / VOR mode.
Prediction for reading out the data of the corresponding mode (VOR/ILS) is the
switching over to display2.
Example:
command from the controller
response of the VOR/ILS Analyzer EVS200
query:
AM8 <CR>
am8 <CR>
e.g.
10.0% <CR/LF>
e.g.
10.0% <CR/LF>