Vor-mode – Procom VOR/ILS Analyzer EVS200 User Manual

Page 53

Advertising
background image

VOR/ILS Analyzer EVS200

0796.1800.02

E-8

44

Chapter 3: Operation

VOR-Mode

Command BE

command

funcion

BE

<Bearing Angle>

bearing angle query

The command BE reads the actual bearing angle (indicated in degree) in the VOR mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

BE <CR>
be <CR>

e.g.

299.97DEG <CR/LF>

e.g.

299.97DEG <CR/LF>

Command AM0

command

function

AM0

<AM-modulation depth (30 Hz)>

query of AM-modulation depth (30 Hz)

The command AM0 reads the actual AM-modulation depth (30 Hz) in the VOR mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AM0 <CR>
am0 <CR>

e.g.

30.1% <CR/LF>

e.g.

30.1% <CR/LF>

Command AM1

command

function

AM1

<AM-modulation depth (9960 Hz)>

query of AM-modulation depth (9960 Hz)

The command AM1 reads the actual AM-modulation depth (9960 Hz) in the VOR mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AM1 <CR>
am1 <CR>

e.g.

30.1% <CR/LF>

e.g.

30.1% <CR/LF>