Specifications, Input characteristics – National Instruments FP-RTD-122 User Manual

Page 11

Advertising
background image

© National Instruments Corp.

11

FP-RTD-122 and cFP-RTD-122

Operate the [c]FP-RTD-122 only at or below Pollution
Degree 2. Pollution Degree 2 means that only nonconductive
pollution occurs in most cases. Occasionally, however, a
temporary conductivity caused by condensation must be
expected.

Refer to the FieldPoint product label for regulatory
certification under hazardous location standards. If the
FieldPoint product is not certified for operation in hazardous
locations, do not operate it in an explosive atmosphere or
where there may be flammable gases or fumes.

Specifications

These specifications are typical for the range –40 to 70°C unless
otherwise noted. Gain error is calculated as a percentage of input
signal value.

Input Characteristics

Number of channels.......................... 8

ADC resolution................................. 16 bits

Type of ADC..................................... Delta-sigma

Input signal ranges (software selectable by channel)

Temperature................................ 73 to 1123 K

–200 to 850 °C
–328 to 1562 °F

Resistance................................... 0 to 400

Ω or 0 to 4,000 Ω

Temperature accuracy (includes 5% matched,
0.5

Ω lead wires—10 m of 22 AWG copper)

Resolution ......................................... 0.016 °C

Resistance accuracy

Offset error, 400

Ω range

15 to 35 °C ........................... 0.03

Ω typ, 0.08 Ω max

–40 to 70 °C......................... 0.08

Ω typ, 0.4 Ω max

Error

Measured Value

15 to 35 °C

–40 to 70 °C

Typical

Maximum

Typical

Maximum

–200 to 150 °C

0.15

0.30

0.40

1.6

150 to 850 °C

0.25

0.50

0.90

3.0