Shock and vibration, Safety, Electromagnetic compatibility – National Instruments FP-RTD-122 User Manual

Page 13

Advertising
background image

© National Instruments Corp.

13

FP-RTD-122 and cFP-RTD-122

Maximum altitude............................. 2,000 m; at higher altitudes

the isolation voltage ratings
must be lowered

Pollution Degree .............................. 2

Shock and Vibration

These specifications apply only to the cFP-RTD-122.
NI recommends Compact FieldPoint if your application is subject
to shock and vibration.

Operating vibration, random
(IEC 60068-2-64).............................. 10–500 Hz, 5 g

rms

Operating vibration, sinusoidal
(IEC 60068-2-6)................................ 10–500 Hz, 5 g

Operating shock
(IEC 60068-2-27).............................. 50 g, 3 ms half sine,

18 shocks at 6 orientations;
30 g, 11 ms half sine,
18 shocks at 6 orientations

Safety

This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement,
control, and laboratory use:

IEC 61010-1, EN 61010-1

UL 3121-1, UL 61010C-1

CAN/CSA C22.2 No. 1010.1

For UL, hazardous location, and other safety certifications, refer to
the product label or to

ni.com

.

Electromagnetic Compatibility

CE, C-Tick, and FCC Part 15 (Class A) Compliant

Emissions.......................................... EN 55011 Class A at 10 m

FCC Part 15A above 1 GHz

Immunity........................................... EN 61326:1997

+ A2:2001,

Table 1

Note

For EMC compliance, operate this device with

shielded cabling.