Physical characteristics, Power requirements, Isolation voltage – National Instruments FP-RTD-122 User Manual

Page 12: Environmental

Advertising
background image

FP-RTD-122 and cFP-RTD-122

12

ni.com

Offset error, 4,000

Ω range

15 to 35 °C ........................... 0.2

Ω typ, 0.7 Ω max

–40 to 70 °C......................... 0.8

Ω typ, 4.0 Ω max

Gain error

15 to 35 °C ........................... 0.01% typ, 0.02% max
–40 to 70 °C......................... 0.07% typ, 0.1% max

Resolution

400

Ω range..........................0.0061 Ω

4,000

Ω range.......................0.061 Ω

Excitation current.............................. 135 ms pulses of 0.25 mA

every 1,080 ms

Input noise

........................................±1 bit peak-to-peak

Input bandwidth ................................ 3 Hz

Update rate........................................ Each channel is updated

every 1.08 s

Physical Characteristics

Indicators .......................................... Green POWER and

READY indicators

Weight

FP-RTD-122 ............................... 140 g (4.8 oz)
cFP-RTD-122 ............................. 110 g (3.7 oz)

Power Requirements

Power from network module ............ 350 mW

Isolation Voltage

Channel-to-channel isolation ............ No isolation between

channels

Transient overvoltage........................ 2,300 V

rms

Environmental

FieldPoint modules are intended for indoor use only. For outdoor
use, FieldPoint modules must be mounted inside a sealed
enclosure.

Operating temperature ...................... –40 to 70 °C

Storage temperature .......................... –55 to 85 °C

Humidity ........................................... 10 to 90% RH,

noncondensing

Advertising