Appendix C

Advanced Hardware Configuration Settings

© National Instruments Corporation

C-3

VXI-USB User Manual

When switch S5 is set so that the VXI-USB receives the SMB CLK10
signal, you have the option to add a 50

Ω termination to the signal by setting

switch S4. S4 is unused when S5 is configured to drive the external CLK
SMB signal.

Table C-1 summarizes the most common configuration types.

Figure C-2 shows the default settings for the CLK10 switches. This
configuration is as follows:

CLK10 is generated from the onboard oscillator (S2).

The CLK10 signal is terminated (S4).

The VXI-USB receives the external CLK10 signal (S5).

The polarity of the CLK10 signal (S6) is irrelevant when the VXI-USB
receives the external CLK10 signal. However, it is configured to be
noninverted when the CLK10 SMB is used as output.

Table C-1. Common CLK10 Routing Configurations

Description

Switches

S2 CLK10

Source

S4

Terminate

S5

Direction

S6

Polarity

Generate internal CLK10 with
onboard oscillator

ONBRD

N/A

IN

N/A

Generate internal CLK10 and drive
to external CLK SMB

ONBRD

N/A

OUT

NON

Generate internal CLK10 and drive
inverted to external CLK SMB

ONBRD

N/A

OUT

INV

Receive external CLK SMB and
drive to the backplane unterminated

SMB

N

IN

N/A

Receive external CLK SMB with
50

Ω termination and drive to the

backplane

SMB

Y

IN

N/A